Published
The principle of the method specified is aspirating the sample into a gas flame of sufficient thermal energy to cause any sodium and potassium present to emit its characteristic radiation and measuring the intensity at a wavelength of 589,0 nm for sodium and 766,5 nm for potassium. Applies to water samples with a mass concentration of sodium and potassium of up to 10 mg/l. For higher concentrations, a smaller test portion is taken for analysis. The lower limits of determination are less than 0,1 mg/l for both sodium and potassium.
PUBLISHED
ISO 9964-3:1993
90.93
Standard confirmed
Dec 12, 2019
Water quality — Determination of sodium and potassium — Part 3: Determination of sodium and potassium by flame emission spectrometry
60.60 Standard published
Only informative sections of projects are publicly available. To view the full content, you will need to members of the committee. If you are a member, please log in to your account by clicking on the "Log in" button.