ISO 7376:2020

Anaesthetic and respiratory equipment — Laryngoscopes for tracheal intubation

Aug 26, 2020

General information

60.60     Aug 26, 2020

ISO

ISO/TC 121/SC 2

International Standard

11.040.10  

English   French  

Buying

Published

Language in which you want to receive the document.

Scope

This document, which is device-specific, specifies requirements for laryngoscopes with non-flexible blades, with internal battery-operated power sources, used for illuminating the larynx during intubation. It also specifies critical dimensions for those handles and laryngoscope blades with interchangeable hook‐on fittings.
It is not applicable to the following:
— flexible laryngoscopes;
— laryngoscopes designed for surgery;
— laryngoscopes powered from mains electricity supply;
— laryngoscopes connected by light‐transmitting cables to external light sources;
— video laryngoscopes designed to work with an external, integral or attached video system.

Life cycle

PREVIOUSLY

Revises
ISO 7376:2009

NOW

PUBLISHED
ISO 7376:2020
60.60 Standard published
Aug 26, 2020

CORRIGENDA / AMENDMENTS

Amended by
ISO 7376:2020/DAmd 1

Preview

Only informative sections of projects are publicly available. To view the full content, you will need to members of the committee. If you are a member, please log in to your account by clicking on the "Log in" button.