This document specifies a method for the measurement of focal spot dimensions within the range of 5 µm to 300 µm of X-ray systems up to and including 225 kV tube voltage. This determination is based on the evaluation of an image with a dedicated focal spot that has been radiographically recorded using an edge and evaluated with a digital method.
The imaging quality and the resolution of X-ray images depend highly on the characteristics of the effective focal spot, in particular the size and the two-dimensional intensity distribution as seen from the detector plane.
For the characterisation of commercial X-ray tube types (i.e. for advertising or trade) the specific maximum values of annex A should be preferred.
IN_DEVELOPMENT
ISO/CD 32543-3
30.99
CD approved for registration as DIS
Jul 2, 2024
Only informative sections of projects are publicly available. To view the full content, you will need to members of the committee. If you are a member, please log in to your account by clicking on the "Log in" button.